FlukeCal 9144-E-P-156 Field Metrology Well (50°C ~ 660°C)

Temperature Range at 23 °C: 50 °C to 660 °C (122 °F to1220 °F)

Display Accuracy: ± 0.35 °C at 50 °C

                             ± 0.35 °C at 420 °C

                             ± 0.5 °C at 660 °C

Stability: ± 0.03 °C at 50 °C

               ± 0.04 °C at 420 °C

               ± 0.05 °C at 660 °C

Axial Uniformity at 40 mm (1.6 in): ± 0.05 °C at 50 °C

                                                       ± 0.35 °C at 420 °C

                                                       ± 0.5 °C at 660 °C

Radial Uniformity: ± 0.02 °C at 50 °C

                             ± 0.05 °C at 420 °C

                             ± 0.10 °C at 660 °C

Hysteresis: 0.1

Immersion (Well) Depth: 150 mm (5.9 in)

Insert OD: 24.4 mm (0.96 in)

Heating Time: 15 min: 50 °C to 660 °C

Cooling Time: 35 min: 660 °C to 50 °C

Resolution: 0.01 °

Display: LCD, °C or °F user-selectable

Size (H x W x D): 290 mm x 185 mm x 295 mm

Weight: 7.7 kg

Power Requirements: 100 V to 115 V (± 10 %), 50/60 Hz, 1400 W 230 V (± 10 %), 50/60 Hz, 1800 W

Computer Interface: RS-232 and 9930 Interface-it control software included