TTI TGR1040 synthesised RF signal generators (10MHz ~ 1000MHz)

Frequency Range: 10MHz to 1000MHz
Setting Resolution: 1kHz by direct keyboard entry, or in user-set increments
of 1kHz to 999·999MHz.
Display Resolution: 1kHz
Accuracy: < ± 1 ppm over temperature range 15°C to 30°C
< ± 2 ppm over temperature range 5°C to 40°C.
Stability: < ± 1ppm/year ageing.
Phase Noise: -110dBc/Hz at 25kHz offset, 500MHz carrier.
Residual FM:
(FM Off)
Equivalent peak deviation in a 300Hz to 3·4kHz
bandwidth: 17Hz at 500MHz carrier.
Output Level Range: –127dBm to +7dBm (0·1µV to 500mV into 50).
Setting Resolution: 0·1dB (or 0·01µV to 1mV) by direct keyboard entry,
or in user-set increments of 0·1dB to 100dB (or
0·01µV to 100mV).
Accuracy: Better than ± 2dBm, except for output levels
<-70dbm at="" 500="" -1000="" mhz="" 3dbm="" br=""> Harmonics: Typically <-25dBc, maximum –20dBc, any carrier frequency,
output level <0dBm.
Non-Harmonic Spurii: <–60dBc at >8kHz offset.
Carrier Leakage: <0.5uv generated="" into="" a="" 50="" load="" by="" a="" 2="" turn="" 25mm="" br=""> loop, at a distance of 25mm from the generator with
the output set to <-10dbm into="" a="" 50="" sealed="" load="" br=""> Output Type: Output impedance 50, Type N connector, Reverse
protection 50V DC.
Output Switch: RF OUT on-off switch with LED showing ON status.
TG
General
Display: 20 character x 4 row backlit alphanumeric LCD
Data Entry: Keyboard selection of frequency, amplitude, etc.;
value entry direct by numeric keys or by rotary control.
Stored Settings: Up to 9 complete instrument set-ups may be stored
and recalled from battery-backed memory.
Mechanical
Size: 130mm (3U) H; 212mm (half-rack) W; 330mm D.
Weight: 4·6 kg. (10 lb)
Power Requirements
100V or 110V - 120V or 220V - 240V, all ±10% 50/60Hz,
adjustable internally; 30VA max. Installation Category II.
Temperature & Environmental
Operating Range: +5°C to 40°C, 20-80% RH.
Storage Range: –20°C to + 60°C.
Environmental: Indoor use at altitudes up to 2000m,
Pollution Degree 2.
Safety & EMC
Safety: Complies with EN61010-1.
EMC: Complies with EN61326.