For full functionality of this site it is necessary to enable JavaScript.

ESPEC PH−402 Temperature Chamber (20~200℃, 1000 L)

System: Forced hot-air circulation / ventilation system

Performance: 

     Temperature range: Ambient temp. +20℃ to +200℃

     Temperature heat-up time: Ambient temp. to +200℃ within 60 min.

Inside dimensions W×H×Dmm (in): 1000×1000×1000 (39.4×39.4×39.4)

Outside dimensions W×H×Dmm (in): 1730×1480×1275 (68.1×58.3×50.2)

Capacity (L): 1000

Weight (kg): 430

Allowable ambient conditions: Temperature: 0 to +40℃; Humidity: to 75%rh

Utility requirements:

Power supply: 200 / 220V AC 3φ 3W 50/60Hz,

                      380V AC 3φ 4W 50Hz

Max. power consumption (kVA): 6.5

INCLUDED ACCESSORIES

● Shelf (stainless steel wire for type102・202): 2 

            (stainless steel punched plate for type 302・402): 2

● Shelf bracket (stainless steel): 2 sets (4)

● Cartridge fuse: 2

● User's manual : 1 set