Working Environment: 

Power supply: 220V AC ±10% 50Hz

Computer: A laptop or a desktop operated with Windows 7 or above and equipped with USB or RS 232 interface

Temperature: 5℃~30℃

Humidity: ≤85%

Technical Parameters: 

Analysis efficiency: 3-4 min each

Weighing accuracy: ≤0.1mg

Detection range: 0.1-500mg nitrogen

Detection recovery rate: ≥99.5%

Relative standard deviation (RSD): ≤0.5% (150mg standard 10% nitrogen)

LOD: 0.01mg nitrogen

Sample weight: Solid ≤1g, liquid ≤1mL

Operating system: Windows 7 and above

Sampler: D200:1 disc x 120 1 disc x 40 positions D50:1 disc x 60 positions

Detector: TCD

TCD accuracy: ≤0.01℃

Carrier gas: CO2

Furnace temperature: 1200℃ (MAX)

Carrier gas purity: 0.99999

Oxygen purity: 0.99999

Internal module communication mode: RS-485 bus

Peripheral interface: RS232, USB

Correction method: Linear or nonlinear calibration, custom calibration curve

Rated power: 2000W

Dimensions (L×W×H): 735mm×560mm×560mm

Net weight: 80kg